In high-performance liquid chromatography (HPLC) and ultra-performance liquid chromatography (UPLC), achieving sharp, symmetric peaks requires seamless compatibility between your sample diluent and mobile phase. The Chromasir® Fidelis-Peak® Solvent Effect Eliminator is a packing-free, finger-tight pre-column device engineered to eliminate peak distortion at the source while protecting your column from particulate contamination.
Engineering Design Principles: Micro-Turbulence Pre-Exchange
The Fidelis-Peak® Solvent Effect Eliminator is a passive microfluidic mixing device installed upstream of the chromatographic column that uses an internal microchannel and cavity structure to generate micro-turbulence, rapidly homogenizing the injected sample solvent with the equilibrated mobile phase before the analyte enters the column bed.
Three-Stage Operating Mechanism
1. Mobile Phase Equilibrium: Installed at the front end of the chromatographic column, the internal cavity of the eliminator becomes completely filled with the mobile phase once the chromatographic system reaches equilibrium.
2. Microchannel Injection & Micro-Turbulence: During sample injection, the sample solution passes through precision microchannels into the small inner cavity. The engineered cavity geometry induces micro-turbulence, enabling immediate and homogeneous mixing between the strong sample solvent and the mobile phase retained inside the cavity.
3. Liquid-Phase Pre-Exchange: This rapid mixing pre-exchanges the liquid-phase environment of the sample, thoroughly resolving mismatches in elution strength, polarity, pH, or organic ratio before the sample enters the chromatographic column.
Technical Advantages Over Conventional Solutions
- • Packing-Free & Restriction-Free: Unlike traditional pre-columns that contain stationary phase packing materials, the packing-free design imposes zero restrictions on chromatographic column selection and avoids stationary-phase degradation.
- • High-Pressure Finger-Tight Design: Withstands operating pressures up to 1000 bar via convenient finger-tight fittings, ensuring leak-free operation across both standard HPLC and UHPLC/UPLC platforms.
- • 316L Stainless Steel & PEEK Core: Built with corrosion-resistant materials and 316L stainless steel as core components for exceptional chemical resistance and an ultra-long operational lifespan.
- • Dual Function as Inline Filter: The integrated internal filter traps solid particles, protecting downstream guard columns and analytical columns from frit clogging.
How to Select the Right Model: HPLC vs. UPLC
Chromasir® provides two dedicated internal volume configurations tailored to system flow rates and dispersion requirements:
| Parameter | 25μL Fidelis-Peak® Solvent Effect Eliminator | 9μL Fidelis-Peak® Solvent Effect Eliminator |
| Part Number | AB1405020 | UP1405007 |
| Outer Dimension | OD: 14mm | OD: 14mm |
| Max Pressure | Finger-tight, 1000 bar | Finger-tight, 1000 bar |
| Wetted Materials | 316L Stainless Steel, PEEK | 316L Stainless Steel, PEEK |
| Target System | HPLC Systems (Standard analytical columns) | UPLC / UHPLC Systems (Low-dispersion columns) |
Step-by-Step Installation Instructions
Proper installation—especially complete air purging—is critical to achieving optimal solvent effect elimination:
1. Position Upstream: Install the Fidelis-Peak® Solvent Effect Eliminator upstream of the chromatographic column (between the injector/mixer and the column inlet).
2. Expel Internal Air Before Column Connection: Before connecting the outlet to your column, hold the eliminator with its outlet pointing upward and flush it with 100% acetonitrile or 100% methanol at a flow rate of 1 mL/min for approximately 2 minutes to thoroughly remove internal air bubbles.
3. Align Flow Direction: After expelling air, connect the eliminator to the inlet of the chromatographic column, ensuring the arrow on the housing aligns with the mobile phase flow direction. (Note: If a guard column is already installed, connect the Fidelis-Peak® Solvent Effect Eliminator before the guard column.)
4. Maintain Consistent Temperature: Place the eliminator alongside the analytical column inside the column oven to maintain a consistent temperature and achieve superior solvent effect elimination.
Operation, Maintenance & Troubleshooting Guidelines
To maximize component lifespan and maintain peak performance, follow these five operational best practices:
1. Routine Use: Because of its packing-free and corrosion-resistant design, conventional chromatographic conditions will not affect its service life, and no special daily maintenance is required.
2. Storage: The device can remain connected and be stored together with the chromatographic column after analysis.
3. Clearing Clogs (High Pressure): If the internal filter traps particles and system pressure rises, disconnect the column and backflush the eliminator at a flow rate of 2 to 3 mL/min.
4. Restoring Separation Performance (Air Entrapment): If column pressure is normal but separation performance deteriorates, air bubbles may have entered the cavity. Position the outlet upward, flush with pure acetonitrile or methanol at 1 mL/min for approximately 1 minute, and then resume operation.
5. Housing Integrity: Do not disassemble the housing of the Fidelis-Peak® Solvent Effect Eliminator unless necessary, as disassembly may impair its internal sealing and microfluidic performance.
Frequently Asked Questions (FAQ)
Where should the Fidelis-Peak® Solvent Effect Eliminator be installed if I use a guard column?
If your chromatographic system includes a guard column, always install the Fidelis-Peak® Solvent Effect Eliminator upstream (before) the guard column, aligning the arrow on the body with the mobile phase flow direction.
What should I do if column pressure increases while using the eliminator?
An increase in pressure typically indicates that the internal filter has trapped solid particulates. Simply disconnect the eliminator from the column and backflush it at a flow rate of 2 or 3 mL/min to clear the trapped debris.
Why is it necessary to flush with 100% methanol or acetonitrile with the outlet upward before first use?
Flushing vertically with 100% organic solvent (acetonitrile or methanol) at 1 mL/min for 2 minutes expels residual air pockets from the microchannels and inner cavity, ensuring homogeneous micro-turbulence mixing during sample injection.
Post time: Nov-10-2026